PHD-12HZ-*11
Overview
| Input signal types and measurement range | ||||
| Code | RTD model | Sukat ning sukad | Minimum range | Katutuan ning pamagbayu |
| 1 | G53 | -50~150°C | 20°C | 0.2°C/0.1% |
| 2 | G50 | -50~150°C | 20°C | 0.2°C/0.1% |
| 4 | Pt100 | -200~850°C | 20°C | 0.2°C/0.1% |
| 6 | Pt1000 | -200~850°C | 20°C | 0.2°C/0.1% |
| 7 | Ni1000 | -60~250°C | 20°C | 0.2°C/0.1% |
Main technical parameters | |
Bilang da reng kanal: | metung a input ampo adwa output |
Suplay ning kuryenti: |
|
boltahe ning suplay ning kuryenti: | 20~35V DC, power consumption a manga 1.8W (24V DC, output 20mA) |
Proteksyun keng kuryenti: | power reverse protection |
Hazardous side input: |
|
Input Signal: | 2-wire o 3-wire RTD signal (lawen ya ing "Input Signal Type and Range Table") |
Signal range: | Katumbas ning sukad ning RTDs. |
Sukat ning sukad: | ing gagamit para gawan ne ing sarili nang configuration neng mag-order ya, makasulat ya keng end number o aliwang instruksiun |
Safety side output: |
|
Output signal: | 4~20mA |
Allowed Output Load Capacity: | 0~500Ω (Customizable) |
LED indicator: | Green ⸺Power indicator; L1 light on para keng low range alarm; L2 light on para keng high range alarm. |
Output accuracy: | lawan me ing "Input signal type and range table". |
Oras ning pakibat: | 300ms anggang 90% ning tauling ulaga |
Temperature Drift: | 0.005%FS/°C |
Parameters ning temperatura: | operating temperature -20°C~+60°C; temperatura ning pamitipun -40°C~+80°C |
Angin relative humidity: | 10%~95%RH alang condensation |
Lakas ning insulasyon: | intrinsically safe end ampo non-intrinsically safe end (≥3000V AC/min); pilatan ning suplay ning kuryenti ampo ing e ligtas (≥1500V AC/min) |
Insulation resistance: | ≥100MΩ (input/output/power supply pilatan) |
Electromagnetic compatibility: | agpang keng IEC 61326-1 (GB/T 18268), IEC 61326-3-1 |
Ing oras para keng kabiguan: | 100000 oras |
Kailangan keng kawad: | cross-section ≥ 0.5mm2; sikanan ning insulasyon ≥ 500V |
Applicable field devices: | 2-wire o 3-wire RTD G53, Cu50, Pt100, Pt1000, Ni1000 |
Lugar ning pamag-install: | maka-install keng safe zone, malyari yang ikonekta keng pekamatas keng Zone 0, IIC, Zone 20, IIIC a delikadung lugal ning intrinsically safe instrumentation |
Sertipikasyun ning kaligtasan | |
Certification body: | National Explosion-proof Electrical Product Quality Supervision and Testing Center CQST |
Certification standard: | GB/T 3836.1-2021, GB/T 3836.4-2021 |
Symbol a e makasabog: | [Exia Ga]IIC, [Exia Da]IIC |
Certified parameters (pilatan da reng terminal 1-2): | Um: 250V AC/DC, Uo=8.4V DC, Io=31mA, |